Total Hemispherical Emittance Measurements

The ET-100 measures directional reflectance from 1.5 to 21 μm and, based on those values, calculates directional and total hemispherical emissivity. The ET100 emissometer conforms to ASTM E408, the standard test method for the determination of emittance using a portable instrument. In-band reflectance data for six discrete bands.

In-Band Reflectance & Emittance

1.5-2.0, 2.0-3.5, 3.0-4.0, 4.0-5.0, 5.0-10.5, 10.5-21.

ASTM Compliant

Compliant with ASTM E408.

Save Time

90 sec warm up, no equilibration
between measurements.

Easy to Use

Simply press port against sample and pull trigger.

Dual Angles of Incidence

Near normal 20° and near grazing 60° incident angles.

Room Temperature Samples

Calculate emissivity without heating sample.

ET100 Applications

 410-VIS-IR for characterizing materials for:

IR signature management
Low observable coatings
Radiative heat transfer
Emissivity for thermal modeling
Thermal camera calibration
Wafer fab hardware emissivity

Intuitive Software

The ET100 touchscreen interface provides:

  • Temperature selection
  • Guided calibration
  • Sample labeling
  • Measurement data display
  • Sample type selection for emittance calculation

More Information

Measured Parameter

Directional hemispherical reflectance (DHR)

Method

Integrated total reflectance in a band for a given angle of incidence

Measured Value

Absolute reflectance (0-1)

Calculated Value

Directional thermal emissivity at 20º, directional thermal emissivity at 60º,
hemispherical thermal emissivity

Wavelength Bands (microns)

1.5-2.0, 2.0-3.5, 3.0-4.0, 4.0-5.0, 5.0-10.5, 10.5-21

Angle of Incidence

20º & 60º from normal incidence

Calibration Coupon

Specular Gold

Accuracy

+/- .03

Repeatability

± .005 units

Beam Spot Size

0.50 inches

Measurement Time

10 sec

Sample Size & Geometry

Flat: ≥ 0.5 in. diameter

Curved: 6 in. convex; 12 in. concave

Warm Up Time

90 seconds

Time Between Measurements

2 seconds

Sample Temperature

Ambient or heated/cooled to 0 – 100º C

Operating Temp

0º to 40º C

Storage Temp

-25º to 70ºC

Run Time

2 hours on one battery. Battery easily replaced with continuous operation after battery replacement.

Power Source

Rechargeable battery (standard environmentally friendly NiMH)

Battery Recharge Time

1 hour

IR Source

Kanthal filament operated at about 1,000ºC

Weight

4.7 lb

Form Factor

H 11.54”, L 9.04”, W 3.27” (29.31 cm x 22.96 cm x 9.44 cm)

Operator Interface

LCD graphics screen, 1/4 VGA, touch screen, software buttons; trigger switch in handle

Software Applications

Pass/fail can be incorporated, user set values

Diagnostics

On screen status and signals monitor. Signal values stored with data. Raw data collection and display.

Data Format

Data files can be opened and post processed with Excel or a text processor

Data Storage

Removable SanDisk (SD) card

ASTM Standard

E903

Export Control

ECCN #3A999.F

Standard ET100 Components

  • Command Module with safety strap
  • ET100 Measurement Head with port cap
  • Specular Gold Calibration coupon in storage container (Non-NIST traceable)
  • SecureDigital card reader on a USB cable
  • SecureDigital card for data storage, 4 GB
  • ET100 operating software
  • 12 Volt battery cartridge – 2 batteries total
  • Battery charger
  • Pelican Case shipping/storage container
  • Accessories container
  • 1-year warranty
  • Worldwide support

Optional ET100 Components

  • Benchtop Remote Control Unit – 220VAC
  • Benchtop Remote Control Unit – 120VAC
  • Specular Gold Calibration Coupon (NIST Traceable)
  • ET100 Reflectometer Maintenance and Calibration Plan (Non-NIST)
  • ET100 Reflectometer Maintenance and Calibration Plan (NIST)
  • ET100 Extended Warranty
  • SD Card for Extra Data Storage
  • Handheld Command Module – 220VAC

Related Products

ET10 Emissometer

410-VIS-IR

Emissivity Measurements