Total Hemispherical Emittance Measurements
The ET-100 measures directional reflectance from 1.5 to 21 μm and, based on those values, calculates directional and total hemispherical emissivity. The ET100 emissometer conforms to ASTM E408, the standard test method for the determination of emittance using a portable instrument. In-band reflectance data for six discrete bands.
ET100 Applications
410-VIS-IR for characterizing materials for:
Intuitive Software
The ET100 touchscreen interface provides:
- Temperature selection
- Guided calibration
- Sample labeling
- Measurement data display
- Sample type selection for emittance calculation
More Information
Measured Parameter |
Directional hemispherical reflectance (DHR) |
Method |
Integrated total reflectance in a band for a given angle of incidence |
Measured Value |
Absolute reflectance (0-1) |
Calculated Value |
Directional thermal emissivity at 20º, directional thermal emissivity at 60º, |
Wavelength Bands (microns) |
1.5-2.0, 2.0-3.5, 3.0-4.0, 4.0-5.0, 5.0-10.5, 10.5-21 |
Angle of Incidence |
20º & 60º from normal incidence |
Calibration Coupon |
Specular Gold |
Accuracy |
+/- .03 |
Repeatability |
± .005 units |
Beam Spot Size |
0.50 inches |
Measurement Time |
10 sec |
Sample Size & Geometry |
Flat: ≥ 0.5 in. diameter Curved: 6 in. convex; 12 in. concave |
Warm Up Time |
90 seconds |
Time Between Measurements |
2 seconds |
Sample Temperature |
Ambient or heated/cooled to 0 – 100º C |
Operating Temp |
0º to 40º C |
Storage Temp |
-25º to 70ºC |
Run Time |
2 hours on one battery. Battery easily replaced with continuous operation after battery replacement. |
Power Source |
Rechargeable battery (standard environmentally friendly NiMH) |
Battery Recharge Time |
1 hour |
IR Source |
Kanthal filament operated at about 1,000ºC |
Weight |
4.7 lb |
Form Factor |
H 11.54”, L 9.04”, W 3.27” (29.31 cm x 22.96 cm x 9.44 cm) |
Operator Interface |
LCD graphics screen, 1/4 VGA, touch screen, software buttons; trigger switch in handle |
Software Applications |
Pass/fail can be incorporated, user set values |
Diagnostics |
On screen status and signals monitor. Signal values stored with data. Raw data collection and display. |
Data Format |
Data files can be opened and post processed with Excel or a text processor |
Data Storage |
Removable SanDisk (SD) card |
ASTM Standard |
E903 |
Export Control |
ECCN #3A999.F |
Standard ET100 Components
- Command Module with safety strap
- ET100 Measurement Head with port cap
- Specular Gold Calibration coupon in storage container (Non-NIST traceable)
- SecureDigital card reader on a USB cable
- SecureDigital card for data storage, 4 GB
- ET100 operating software
- 12 Volt battery cartridge – 2 batteries total
- Battery charger
- Pelican Case shipping/storage container
- Accessories container
- 1-year warranty
- Worldwide support
Optional ET100 Components
- Benchtop Remote Control Unit – 220VAC
- Benchtop Remote Control Unit – 120VAC
- Specular Gold Calibration Coupon (NIST Traceable)
- ET100 Reflectometer Maintenance and Calibration Plan (Non-NIST)
- ET100 Reflectometer Maintenance and Calibration Plan (NIST)
- ET100 Extended Warranty
- SD Card for Extra Data Storage
- Handheld Command Module – 220VAC